Vernier Spectrometer
|
Ocean Optics Red Tide Spectrometer
|
Ocean Optics Red Tide UV-VIS Spectrometer
|
| Ocean Optics Model |
|
USB650 and USB-ISS-VIS |
USB650-UV-VIS |
| User Manual |
Download |
Download |
Download |
| Physical |
| Dimensions (mm) |
89.1 × 63.3 × 34.4 |
| Weight (g) |
190 |
| Detector |
| Type |
Linear Silicon CCD Array |
| Pixels |
325 enabled pixels |
650 enabled pixels |
| Pixel Size (µm) |
14 x 200 |
| Pixel Well Depth |
~62,500 |
| Sensitivity (photons/count @ 400 nm) |
75 |
| Optical Bench |
| Design |
f/4, asymmetrical crossed Czerny-Turner |
| Focal Length (mm) |
42 (input); 68 (output) |
| Entrance Aperture |
25 µm wide slit |
| Fiber Optic Connector |
SMA 905 |
| Grating |
Grating #2 |
Grating #1 |
| OFLV Filter |
|
OFLV-200-850 |
| UV enhanced window |
|
Yes, UV4 |
| Spectroscopic |
| Wavelength Range (nm) |
350-1000 |
200-850 |
| Signal-to-Noise Ratio |
250:1 (at full signal) |
| A/D Resolution (bit) |
12 |
| Dark Noise |
3.2 RMS counts |
| Dynamic Range |
2 × 108; 1300:1 for a single acquisition |
| Integration Time |
3 ms to 65 s (15 s typical maximum) |
| Stray Light |
<0.05% @ 600 nm; <0.10% @ 435 nm |
| Corrected Linearity |
>99.8% |
| Electronics |
| Power consumption |
|
90 mA @ 5 VDC |
| Data transfer speed |
|
Full scans to memory every 13 ms with USB 2.0 port, 18 ms with USB 1.1 port |
| Breakout Box Compatible |
|
Yes, Requires Adapter |
| Trigger Modes |
|
5 |
| Connector |
|
10-pin connector |