Vernier Tech Info Library TIL #1415
EKG Sensor Detail InformationEKG Sensor, order code EKG-BTA, $147.00
Specifications and User Guide
TROUBLESHOOTING
-Primary Test: Confirm that the electrode patches are fresh and connected to the sensor properly. A drawing on the sensor box shows the proper connections. Results vary form person to person. If an individual always has trouble getting a clean signal, try alcohol swabs before affixing the electrodes. The mostly cause of problems with this sensor is a poor connection, either between the tabs and the skin or between the clips and the tabs. Note that electrode tabs do not work as well if they are old or have been allowed to dry out.
-Secondary Test: Make sure that the person connected to the sensor is seated or lying down and that they are relaxed.
RELATED TIL ENTRIES
TIL 2036: Why is my EKG trace noisy? Why doesn’t my EKG trace look like the trace in the book?
TIL 2037: Can I get Logger Pro to display Heart Rate from EKG data
TIL 1740: Could a cell phone affect EKG data?
SPECIFICATIONS
-new order code: EKG-BTA
-old order code: EKG-DIN
-Offset: ~1.00 V (+/-0.3 V)
-Gain: 1 mV body potential/1 V sensor output
CALIBRATION
-Calibrate? No. The sensor is set to the stored calibration before shipping. No calibration process is given for this sensor.
RELATED VERNIER PRODUCTS
-EKG Electrodes, order code ELEC, $12.00 (one pkg)
Created by: jrandall on May 07 2005
Last updated by: jhopkins on August 17 2009