SUPPORTED CALCULATOR APPS

  • EasyData – for TI-83 Plus and TI-84 Plus calculators and TI-SmartView emulator software
  • DataQuest – for TI-Nspire handhelds and computer software

TROUBLESHOOTING

The LED on EasyTemp is a useful troubleshooting tool. When working correctly, the LED on EasyTemp is green. If your LED is not green, see: The LED on my Go! device or LabQuest Mini is not green. What does that mean?

SPECIFICATIONS

– Range: -20°C to 115°C
– Maximum temperature with : 150°C
– Resolution: 0.07°C
– Accuracy: ± 0.5°C
– Response time: 4 s (to 90% of full reading in water)

CALIBRATION

Calibrate? EasyTemp is very nicely calibrated when shipped. Most never need to calibrate. EasyTemp uses a linear calibration equation and can be calibrated if required. See the data collection app user manual for details on how to calibrate a sensor.

RELATED VERNIER PRODUCTS

ADDITIONAL INFORMATION