Yes – with limitations. Vernier spectrometers can be used to explore blackbody radiation qualitatively, such as observing general trends in emission spectra. However, they are not calibrated for absolute intensity, so accurate quantitative comparisons across wavelengths or matching to Planck’s Law are not possible.

⚠️ Understanding Intensity Mode and Detector Response in Vernier Spectrometers
The spectral response of Vernier spectrometers is not flat—meaning a reading of 0.5 intensity at 400 nm does not represent the same absolute light intensity as a reading of 0.5 at 600 nm. This is because the detector, optics, and system components all respond differently across the visible and UV spectrum.

🧪 Why This Matters: Intensity Mode vs. Absorbance Mode
In Absorbance Mode, this non-flat response doesn’t matter. The spectrometer calibrates using a blank (typically a clear cuvette with solvent), so any wavelength-dependent variation is automatically accounted for.
In Intensity Mode, however, no calibration is applied. The raw readings you see reflect not just the intensity of the light source but also the variable sensitivity of the detector system.

🔍 Can I Use Intensity Mode for Comparing Light Sources?
Yes—with limitations:

  • You can make relative comparisons of intensity between samples measured at the same or similar wavelengths.
  • Do not compare intensity values across very different wavelengths, as detector response varies significantly across the spectrum.

🌡️ What About Measuring Blackbody Radiation?
If you’re using the spectrometer to measure a blackbody radiator (like an incandescent bulb or heated object), the non-flat response of the detector will distort the shape of the curve. The peak of the blackbody spectrum may appear shifted or suppressed in regions where the detector or optics have lower sensitivity.

Important: Vernier spectrometers are not calibrated to report true irradiance or spectral power density. They are not suited for quantitatively matching experimental blackbody curves to theoretical models like Planck’s Law without applying complex post-processing corrections.

⚙️ Why Isn’t Intensity Response Calibrated?
Calibrating a spectrometer for flat, wavelength-corrected intensity is extremely complex. Even small changes—like removing and reinserting a fiber optic cable—can invalidate such a calibration.

System-wide response varies due to:

  • Mirrors and gratings
  • Fiber optics
  • Manufacturing differences in detectors

Because of this, Vernier spectrometers are designed for relative intensity work, not calibrated irradiance measurements.

Wavelength Accuracy Is Still Reliable
While intensity values are uncalibrated, wavelength calibration is highly accurate. The position of spectral peaks is reliable and does not require further adjustment.

📘 Learn More:
When should I use the Intensity Correction for my Vernier-branded Spectrometer?