Troubleshooting

  • Primary Test: In air, the Soil Moisture Sensor should read somewhere between -5% and -50%. Grab the sensing portion with your hand, and the reading should go up to somewhere between 10% and 50%.

Additional Troubleshooting

Specifications

  • Range: 0 to 45% volumetric water content in soil
    (capable of 0 to 100% VWC with alternate calibration)
  • Accuracy: ±4% typical
  • Typical Resolution: 0.1%
  • Power: 3 mA @ 5VDC
  • Operating temperature: –40°C to +60°C
  • Dimensions: 8.9 cm × 1.8 cm × 0.7 cm (active sensor length 5 cm)
  • Default calibration coefficients: Slope = 108%/V; Y-intercept = –42%

Calibration

Calibrate? Usually, no. Possible exceptions: you are testing soil with high concentrations of salt, sand, or organic matter; very precise readings are required. See detailed calibration instruction in the user manual. For additional information on sensor calibration, see How do I calibrate my sensor?