My spectrometer is connected to a computer and I get the error message, "Could not collect values from device," during calibration.

This can happen when Logger Pro can't tell the difference between the dark and light sample during calibration. If you are using a standard, water blank (nothing opaque), then the light source is probably not operating correctly. This can be caused by a dead bulb, an underpowered USB port, or when the spectrometer has lost its calibration. Try removing non-essential USB devices and connecting the spectrometer directly to the computer--not through a USB hub. Try every USB port on the computer just in case some provide less power than others. If none work, try it on another computer. If this still doesn't work, the bulb may be burned out or the detector may have lost calibration, or more than one of the above.

To check the light source See TIL 2416: How do I check the lamp output of my spectrometer?

If you are using SpectroVis Plus, please use Logger Pro or LabQuest to locate the serial number of the device. If the serial number is 000000 then the calibration has been lost on the device and it needs to be sent in to be recalibrated.

Logger Pro
The serial number is located in the spectrometer dialog box.
2408_LP_SVIS.PNG

LabQuest
The serial number can be found by tapping on Mode: Full Spectrum.
2408_LQ_SVIS.png