Troubleshooting
- Primary Test:
- Connect the Constant Current System to the interface.
- Connect the AC power adapter. (Included, but the same as LabQuest® Power Supply (LQ3-PS))
- Start the data-collection software.
- Be sure the dial is turned counterclockwise to the minimum. Connect the red and black clips together.
- Turn the dial clockwise. The current reading should increase to 0.6 A when the dial is turned to the maximum.
Additional Troubleshooting
- Do you have updated instructions for Advanced Chemistry with Vernier Experiment 21, Electroplating?
- Do you have a power supply for electrolysis?
Specifications
- Range: 0-0.6 A
- Typical Resolution: 0.09 mA
- Linearity: 0.1%
- Supply voltage: 5 VDC
- Current in amperes:
⚬ Slope: 0.145 A/V
⚬ Intercept: 0 A - power supply: LabQuest® Power Supply (LQ3-PS)
Calibration
Calibrate? No. The sensor is set to the stored calibration before shipping. To perform your own calibration, refer to the user guide.