In the situation that your Go Direct® Optical Dissolved Oxygen Probe (
To do a reset, follow the instructions below.
- Fill the storage bottle to the top of the sponge with distilled water.
- Insert the probe into the storage bottle and make sure it is positioned vertically so that the tip is not in contact with the sponge or water.
- Start the software and identify your probe, if necessary.
- Keep the probe in a vertical position for a minimum of 60 seconds.
- Click or tap the sensor meter for the channel you want to reset (DO Concentration or DO Saturation).
- Choose calibrate, and follow the instructions on the screen.
If resetting the calibration does not contact Vernier Support at support@vernier.com or 1-888-837-6437 for a return merchandise authorization for your sensor to be evaluated.