Troubleshooting

  • Primary Test:
    1. Connect the Spirometer to a Vernier interface (e.g., LabQuest, LabQuest Mini, or Go!Link).
    2. Launch the Data Collection software.
    3. Hold the sensor upright and still, then “Zero” the sensor reading. Once zeroed, confirm that the flow rate is zero or very nearly so.
    4. Connect a bacterial filter and a mouthpiece to the side of the sensor marked “Inlet.”
    5. Start data collection and exhale into the sensor. Confirm that the flow rate and lung volume are reasonable values for the person exhaling into the sensor.

Additional Troubleshooting

Specifications

  • Flow rate: ±10 L/s
  • Dead Space: 93 mL
  • Nominal Output: 128 mV/[L/s]
  • Detachable Flow Head
    ⚬ Dimensions: 80.5 mm (diameter) × 101.5 mm (length)
    ⚬ Mass: 80 g
    ⚬ Construction: Clear acrylic plastic
  • Handle
    ⚬ Dimensions: 127 mm × 23 mm × 35 mm
    ⚬ Mass: 85 g
    ⚬ Construction: Black ABS plastic
  • Cable length: 1.5 meters
  • Stored calibration coefficients: Slope = 7.8 (L/s)/V; Intercept = –17.55 L/s

Calibration

Calibrate? No. The sensor is set to match the stored calibration before shipping. No calibration process is recommended for this sensor.