Aug 12, 2019 ... -Calibrate? Yes. Conduct a two-point calibration, using the standard solutions that are shipped with the sensor. First point: The ISE should be ...
Conduct a two-point calibration, using the standard solutions that are shipped with the sensor. First point: The ISE should be soaking in the High Standard.
Jan 25, 2019 ... ... ISE is in the High and Low standards. These values are ... NH4 1 mg/L Standard NH4 ISE Solution (NH4-LST) NH4 100 mg/L Standard ...
Yes, it is possible to use a Vernier Ion-Selective Electrode (ISE) with an Arduino microcontroller. However, we do not recommend this combination unless you are...
Standard cleaning is best done with DI water. If sterilization is needed, use ethyl alcohol (typically >90%) or isopropyl alcohol (typically 60-70%) and swirl in...
We do not recommend using our sensors for long-term data collection. They are intended for use in a classroom during a lab period. Alternative suppliers...
Specifications and User Guide TROUBLESHOOTING -Primary Test: Check the raw voltages while calibrating in the standards. The high (1000 mg/L) should be around 2.7 volts....