Since the PH-BTA has more exposed glass, it is typically easier to break. As an additional note, if the Flat pH electrode is broken, the FPH-BNC can be ...
-Secondary Test: Take readings in fresh buffers, NOT distilled water! If no buffers are available, try it in vinegar (~ pH 2.5-3.5) and ammonia (~ pH 10.5-11.5)Â ...
Nov 7, 2019 ... Tris-Compatible Flat pH Sensor (FPH-BTA) (the replacement electrode is Flat pH Electrode BNC (FPH-BNC)) Go Direct Tris-Compatible Flat pHÂ ...
Jul 13, 2017 ... 4. Once the voltage reading has stabilized, tap Add in the box under the appropriate buffer pH. Note: If you are using a buffer not shown, tap ...
Dec 8, 2015 ... The pH Sensor (PH-BTA or PH-BNC) will have difficulty measuring the pH of the following liquids: distilled water, water with a low ion ...
If you calibrate a PH-BNC electrode with an Electrode Amplifier (EA-BTA), typical potential readings are -pH 4 buffer solution: 1.56 volts -pH 7 buffer solution:...
Generally, rinsing the electrode with distilled water after using should suffice. If the glass bulb looks dirty, you can rinse the electrode with warm water...
We carry pH/ORP Storage Solution with buffer preservative (mold inhibitor) added. Each bottle contains 500 mL of pH/ORP Storage Solution. We sell this as ....