Feb 6, 2015 ... Primary Test: Check the raw voltages while calibrating in the standards. The high (100 mg/L) should be around 1.6 volts. The low (1 mg/L) should ...
Place the electrode into the Low Standard. Enter the concentration value of ... Standard NO3 ISE Solution (NO3-HST). Published: August 12, 2019. Updated ...
Jan 25, 2019 ... ... ISE is in the High and Low standards. These values are ... NO3 100 mg/L Standard NO3 ISE Solution (NO3-HST). Ca 10 mg/L Standard ...
Nov 10, 2016 ... Standard Cl ISE Solution (CL-LST)—Low Concentration; Standard Cl ISE Solution (CL-HST)—High Concentration; Electrode Storage Bottles (BTL-ES).
Feb 6, 2015 ... Standard Ca ISE Solution (CA-LST)—Low Concentration; Standard Ca ISE Solution (CA-HST)—High Concentration; Electrode Storage Bottles (BTL-ES).
(The green LED will flash to indicate the sensor is connected.) Primary Test: Check the readings in the standards. The high should be around 1000 mg/L. The low ...
May 5, 2015 ... Remove the ISE from the High Standard. ... Replacement Parts. Potassium Replacement Module (K-MOD); Standard Potassium Solution (K-LST)—Low ...
Yes, it is possible to use a Vernier Ion-Selective Electrode (ISE) with an Arduino microcontroller. However, we do not recommend this combination unless you are...
Standard cleaning is best done with DI water. If sterilization is needed, use ethyl alcohol (typically >90%) or isopropyl alcohol (typically 60-70%) and swirl in...
When an Ion Selective Electrode (ISE) is connected to a Go Wireless Link, the attached sensor can be identified incorrectly. This happens when you have...