Feb 6, 2015 ... Primary Test: Check the raw voltages while calibrating in the standards. The high (100 mg/L) should be around 1.6 volts. The low (1 mg/L) should be around 2.4Â ...
Jan 25, 2019 ... ... ISE is in the High and Low standards. These values are ... NO3 100 mg/L Standard NO3 ISE Solution (NO3-HST). Ca 10 mg/L Standard ...
Aug 12, 2019 ... -Calibrate? Yes. Conduct a two-point calibration, using the standard solutions that are shipped with the sensor. First point: The ISE should be ...
Standard cleaning is best done with DI water. If sterilization is needed, use ethyl alcohol (typically >90%) or isopropyl alcohol (typically 60-70%) and swirl in...
Specifications and User Guide TROUBLESHOOTING -Primary Test: Check the raw voltages while calibrating in the standards. The high (1000 mg/L) should be around 2.7 volts....
When an Ion Selective Electrode (ISE) is connected to a Go Wireless Link, the attached sensor can be identified incorrectly. This happens when you have...