If your spectrometer isn’t calibrating or is showing unusually noisy or high absorbance values (often above 3 or blank), the issue may be due to insufficient light reaching the detector. This happens when:
The light source is weak in a certain...
Sometimes the counts from the light striking a pixel is not sufficiently different from the counts when the pixel is in the dark to make a reliable absorbance reading. This can happen either when the light source is weak in that spectral region, or w...
The Conductivity Probe (CON-BTA) is automatically temperature compensated between temperatures of 5 and 35°C. Note that the temperature of a solution is being read by a thermistor that extends into the space between the graphite electrodes. Readings...
Below are some of the sports-related activities we have done using Vernier sensors.
Using a Goniometer to Explore Heart Rate and Cadencehttps://www.vernier.com/vernier-ideas/using-a-goniometer-to-explore-heart-rate-and-cadence/
Analysis of Bar...
Yes. Go Wireless and some Go Direct sensors can be used with Google Science Journal.
For more information on Science Journal, see
https://www.vernier.com/products/software/science-journal/
https://makingscience.withgoogle.com/science-journa...
Force Plate (FP-BTA)
Specifications and User Guide
Which force plate do I have?The original Force Plate (shown here) was revised in November 2023.Original Force Plate Specifications and User Guide
Original Force Plate
...
Supported Go Wireless Sensors
Go Wireless® Heart Rate (GW-HR)
Go Wireless® Exercise Heart Rate (GW-EHR)
Graphical Analysis app: these sensors are fully supported (https://graphicalanalysis.app).Logger Pro or Logger Lit...
To get the most accurate readings from the Voltage Probe, when measuring the half cell potential in a chemistry experiment (such as Lab 28 from CwV) or from a standard dry cell battery, you should zero the probe. There is a specific process for zeroi...
Unlike the LabQuest version of the Conductivity Probe which has three ranges that you change manually with a switch, the Go Direct Conductivity automatically changes range based on the solution you are in. Because of this, the best way to calibrate i...