Please read the user manual for general usage information.

Troubleshooting

  • Primary Test: Are you using LabQuest App v2.7.2, or Spectral Analysis v4.4 (or newer)?
  • Secondary Test: If you are collecting data via USB, you do not need the battery installed or the AC power cable connected. Removing both of these will likely solve most issues.

In general, we recommend removing the Go Direct SpectroVis Plus battery if you are not using the device wirelessly. The AC power adapter is only needed to charge the battery and is not required for operation.

Additional Troubleshooting

MOST COMMON

GENERAL

ACCESSORIES

SOFTWARE

EXPERIMENTS

Specifications

  • Light Sources
    ⚬ Incandescent white bulb (~ 8000 hour lifetime)
    ⚬ LED-based support (~100,000 hour lifetime)
  • Fluorescence: Excitation LEDs centered at 405 and 500 nm
  • Detector: Linear CCD
  • Slit Width: 25 μm
  • Wavelength Range: 380 to 950 nm
  • Wavelength Reporting Interval: ~1 nm (collects 570 values)
  • Optical Resolution (FWHM): 5.0 nm
  • Wavelength Accuracy: ±4.0 nm
  • Photometric Accuracy: ±0.10 A.U.
  • Typical Scan Time: ~2 s
  • Operating Temperature: 15 to 35ºC
  • Dimensions: 15 cm × 9 cm × 4 cm
  • Z Dimension (Beam Height): 8.5 mm
  • Connections
    ⚬ Wireless: Bluetooth® v4.2 (wireless range 30 m unobstructed)
    ⚬ Wired: USB 2.0 full speed
  • Battery: High-capacity, lithium-ion rechargeable battery

Calibration

Calibrate this sensor? Yes, in absorbance and transmittance modes. Calibration when using emissions or fluorescence modes is optional. See calibration instructions below:

  1. Start the calibration.
    ⚬ Logger Pro: Choose Calibrate ▶ Spectrophotometer from the Experiment menu.
    ⚬ LabQuest App: Choose Calibrate ▶ USB: Spectrophotometer from the Sensors menu.
    ⚬ Spectral Analysis: Select to start an Absorbance or Transmittance experiment.
    Note: For best results, allow the Spectrophotometer to warm up for a minimum of five minutes.
  2. Fill a cuvette about 3/4 full with distilled water (or the solvent being used in the experiment) to serve as the blank. After the spectrophotometer has warmed up, place the blank cuvette in the spectrophotometer. Align the cuvette so a clear side of the cuvette is facing the light source.
  3. Follow the instructions in the dialog box to complete the calibration, and then tap OK.